Alarm Messages Savina Version 2.0
1.
Voltage supply is faulty.
2.
A/D converter faulty.
1.
The operating voltages are cyclically compared to defined tolerance limits (see Tabl
error detection is provided with an error counter per voltage to be monitored.
2.
The input value "Voltage faulty (SiSW)" reports an error.
1.
None.
Supply voltages
Tolerance
+15 V
±20%
-15 V
±20%
+24 V
±20%
+5 V reference voltage
±4%