2.
Replacement of wear and tear parts
2.1
Replace diaphragm, complete, 84 13 661, in
the expiratory valve.
Next replacement:___________________
2.2
Replace internal batteries (rechargeable), 18
41 416, (x2) every 2 years
(see also chapter "Maintenance Procedure
Replacing the internal rechargeable
batteries
").
Next replacement:___________________
2.3
Replace batteries (rechargeable), 18 43 303,
(if fitted, x2) in Savina Mobile every 2 years.
Next replacement:___________________
2.4
Replace O2 filter, 84 08 208, and sealing
ring, M 09 257, in valve block, complete,
every 6 years.
(see also chapter "Maintenance Procedure
Replacing filter and sealing ring in O2
compressed gas inlet
").
Next replacement:___________________
2.5
Replace O2 pressure reducer, 84 13 666, in
valve block, complete, every 6 years.
(see also chapter "Maintenance Procedure
O2 pressure reducer
").
Next replacement:___________________
2.6
Replace real-time clock, 18 45 527, on
Control PCB, every 6 years.
(see also chapter "Maintenance Procedure
Replacing the real-time clock on the
Control PCB
").
Next replacement:___________________
Note: After changing the real-time clock re-
enter the date and time.
2.7
Replace the blower unit with the "motor
blower unit" spares set, 84 13 643, after
20000 hours’ operation.
(see also chapter "Maintenance Procedure
Replacing the blower unit
").
*
*
*
*
*
*
*