A
B
C
4.16
Testing the optional central alarm (if available)
Activate alarm, e.g. by pulling power plug.
Test set-up:
Test value:
Contacts "3" and "5" close if an alarm occurs.
Contacts "1" and "3" open.
4.17
Service hours counter and log entries
Switch off Savina.
Switch on Savina and, at the same time, press
and hold the backlighting dim-up/dim-down
key and the "Config" key until the service
mode is activated (see also chapter "Service
Mode
Entering the Internal Service Mode
").
Call up test
0
0
:
Perform "DELETE LOGBOOK" function.
Call up test
3
3
:
Perform "RESET SERVICE HOURS"
function.
5.
Place fully functional unit at user’s/owner’s
disposal.
6.
Confirmation of test
Name:
Date:
Signature:
These steps are regarded as repair work and
are therefore not included in the inspection
service price.
Ohmmeter
Central alarm socket (located on the
rear panel of the unit)
F
*