3.2.4.3
Earth leakage current test
The test is carried out without Savina or other
devices.
Normal condition (N.C.):
I
≤
50
µ
A
Single fault condition (S.F.C.):
Neutral conductor interrupted:
I
≤
100
µ
A
In the following steps the test is repeated, but
with the power plug turned over. This
condition can be created internally with some
types of test devices.
Normal condition (N.C.):
I
≤
50
µ
A
Single fault condition (S.F.C.):
Neutral conductor interrupted:
I
≤
100
µ
A
Continue Test Certificate at test item
4.
.
3
3..3
3
T
Te
es
st
t o
of
f e
elle
ec
cttrriic
ca
al
l s
sa
affe
etty
y tto
o V
VD
DE
E 0
07
75
51
1
3.3.1
Protective earth conductor test
Connect the test probe in sequence to the
following test points:
−
Power pack ground stud
−
Screws on housing
−
Rails, at the sides of Savina (if fitted)
−
Oxygen connection
Test value R
≤
0.3 ohms
P
P
P
P
Tester
Savina
6 V
ohm
L
N
PE
Test probe
PS
Protective earth conductor test
Protective earth conductor test
P