Connect airway temperature sensor (1) to the
socket of the airway temperature sensor of
Savina and to the tester. Connect test probe
to Savina.
Enter serial number of the airway temperature
sensor:
Subsequent measurements may exceed the
first-measured value by max. 50% but must at
the same time be
≤
5000
µ
A.
Initial value: _______
µ
A
Note:
Always enter the initial value in a new PMS.
I
≤
5000
µ
A
3.3.4
Multiple socket outlet on trolley (if fitted)
3.3.4.1
Compare fuses with rated values on the
socket-outlet.
3.3.4.2
Protective earth conductor test
Check all sockets.
Test value R
≤
0.3 ohms
3.3.4.3
Equivalent device leakage current test
The test is carried out without Savina or other
devices.
Subsequent measurements may exceed the
first-measured value by max. 50% but must at
the same time be
≤
100
µ
A.
Initial value: _______
µ
A
Note:
Always enter the initial value in a new PMS.
I
≤
100
µ
A