Dr
ä
ger Medizintechnik GmbH
Error List Savina Version 2.0
60.0011
Inspiratory airway pressure sensor
S4.2: Measured value is outside the
measuring range.
−
Pressure applied to the airway
pressure sensor was outside the range
of
—
10 to +110 mbar.
−
Check connection from solenoid and
airway pressure sensor to the
PCB.
60.0012
Inspiratory airway pressure sensor
S4.2: Deviation between maximum
and minimum value is too small.
−
Check connection from solenoid and
airway pressure sensor to the
Control PCB.
−
Remove stenosis in hose of supply to
airway pressure sensor.
60.0020
Calibration error S 4.3 (not used for
the time being).
−
60.0021
Measured value S 4.3 outside of
measuring range (not used for the time
being).
−
60.0100
Deviation between inspiratory and
expiratory airway pressure is too large
(not used for the time being).
−