Alarm Messages Savina Version 2.0
1.
The ventilator does no longer generate control information of the flow-control unit.
2.
Due to device settings, only very low volumes can be applied if the patient
’
s lungs are hard.
1.
If, in pure controlled modes (IPPV, SIMV), the "ExpFlow start" and "ExpFlow end" info
is missing for more than the set apnea alarm time, this will be regarded as "Fail to Cycle".
1.
After a cold start and 15 s afterwards.
2.
During standby and 15 s afterwards.
3.
During a detected disconnection and for the period T
apnea
afterwards.
4.
An "Airway pressure high" alarm has occurred.
5.
An "Airway pressure low" alarm has occurred.
6.
Hold condition is active (with insp. hold).