Service Mode Savina Version 2.0
Test step 10 is used to display the expiratory pressure "Pawexp" (S4.1) and the inspiratory
pressure "Pawinsp" (S4.2). The maximum permissible pressure is 105 mbar. The externa
pressure measuring device indicates the opening pressure of the safety valve D1 (Target
value = 102 mbar to 110 mbar).
The blower delivers a flow 30 L/min; the expiratory valve is "sealed" with 120 mbar.
Connect (short-circuit) the inspiratory socket and the expiratory socket with a ventilation