Dr
ä
ger Medizintechnik GmbH
Service Mode Savina Version 2.0
Test step 17 is used to display the measured value of the internal A/D converter of the master
processor.
V_FLOW_TEMP
0409 ... 0512 2.00 V ... 2.50 V (at ambient temperature). Tem
of inspiratory gas, measured at TSI
sensor (S1) assembly.
V_TEMP_AIR
0246 ... 0327 1.20 V ... 1.60 V (at ambient temperature). Tem
of inspiratory gas. The temperature
sensor (NTC) is located on the
inspiration block. The measurement
signal is processed on the O2
Diaphragm PCB.
VCC+15V_O2_MESS 0593 ... 0634 2.90 V ... 3.10 V Measurement signal of the +15
voltage of the O2 Diaphragm PCB/O2
sensors.
VCC+15V_MESS
0573 ... 0655 2.80 V ... 3.20 V Measurement signal of the +15
voltage.