Alarm Messages Savina Version 2.0
As described in chapters
FiO2 low
and
FiO2 high
, the measured value FiO2(t) is compared to an
upper and a lower limit value. An alarm is generated if there is a deviation. These two limit values
are generated as follows:
Normal operation (concentrator is not active)
−
FiO2 low = FiO2 target value -8 vol.%
−
FiO2 high = FiO2 target value +8 vol.%
−
Fixed alarm limit = FiO2 low
≥
18 vol.%
Concentrator is active
−
The FiO2 alarm limits set by the user are used.
Remark: Bronchial suction (oxygenation) changes the set FiO2 concentration to 100%. The limit
value is changed at the same time.