Dr
ä
ger Medizintechnik GmbH
Error List Savina Version 2.0
60.0001
Inspiratory airway pressure sensor
S4.1: Measured value is outside the
measuring range.
−
Pressure applied to the airway
pressure sensor was outside the range
of
—
10 to +110 mbar.
−
Check connection from solenoid and
airway pressure sensor to the
Control PCB.
60.0002
Inspiratory airway pressure sensor
S4.1: Deviation between maximum
and minimum value is too small.
−
Check connection from solenoid and
airway pressure sensor to the
Control PCB.
−
Remove stenosis in hose of supply to
airway pressure sensor.
60.0010
Expiratory airway pressure S 4.2:
Calibration error.
−
Check connection from solenoid and
airway pressure sensor to the
PCB.
−
See Service Mode, test 19:
Replace expiratory pressure sensor
S4.2 if test 19 fails. Replace solenoid
V6.2 if test 19 is OK.