Service Mode Savina Version 2.0
Test step 18 is used to display the measured value of the internal A/D converter of the master
processor.
V_FIO2_1
0102 ... 0205 0.50 V ... 1.00 V (if system is free from O2).
Measurement signal of O2 sensor 1.
The measurement signal is processed
on the O2 Diaphragm PCB.
V_O2_PRESSURE
0164 ... 0246 0.80 V ... 1.20 V (O2 depends on supply pressu
disconnect O2 supply). O2 supply
pressure, measured at S5. S5 is located
on the O2 Valve PCB.
V_REF_O2
0593 ... 0634 2.90 V ... 3.10 V Reference voltage for
V_TEMP_AIR
and
V_TEMP_O2
. The reference
voltage is generated on the O2
Diaphragm PCB from the battery
voltage.
V_PAW_REG
0123 ... 0184 0.60 V ... 0.90 V Measurement signal of control
sensor S4.3. The sensor is only
available in the first lot of units.