Repair Information/Error Events Sav ina Version 2.0
For the power pack to charge the internal battery, it is sufficient to increase the battery voltage
shortly above 10 V or 20 V, respectively. This can be done by applying 24 V DC to the batt
Connect Savina to mains power supply and apply the 24 V DC voltage to the connectors as
follows:
Rear view of external voltage supply
Protect the 24 V DC voltage with an external fuse of 15 A or use a voltage supp
current limiter. Set the current limiter to 1 A.
Apply the 24 V DC voltage until the "internal battery" LED comes on (yellow) and stays
on. The LED is located on the front panel of Savina.
I
Left, bottom connector of external DC voltage
supply: Negative pole
Left side of internal battery fuse: Positive pole