Maintenance Procedure Savina Version 2.0
2
To adjust the pressure reducer the valve block must be removed (see "
Replacing filter and sealing
ring in O2 compressed gas inlet
". Follow the instructions given there from "
Switch off the unit.
"
Remove the valve block.
").
•
Connect the O
2
Valve PCB to the Savina.
•
Connect the O
2
pressure supply.
•
Connect pressure gauge, 0 to 6 bar (Use valid test equipment. The illustration shows only a
schematic) to the nebulizer outlet of the valve block.
Pressure reducer test set-up
O
2
Valve PCB connector
O
2
pressure supply
Nebulizer outlet