Dr
ä
ger Medizintechnik GmbH
Function Description Savina Version 2.0
The pneumatic system comprises the following principal components (see also
Fig.9: Pneumatic
diagram
:
−
Drawer unit
−
Valve block
−
Inspiration block
−
Pressure measuring unit
−
Patient system
−
Flow sensors
The following summary describes the functions of the various principal components rather than
giving a detailed description of the components themselves.
The blower draws ambient air through an ambient air filter F1 into a mixing chamber. The blower
compresses the gas to a positive pressure of max. 140 mbar at a flow rate of up to 180 L/mi
blower is run at constant rotation speed - that is, the speed is not varied in the course of a
The inspiratory pressure is regulated by means of the bypass valve. The combination of the blower
and the bypass valve function thus represents a pressure source.
On the inlet and outlet sides of the blower there are mufflers SD1 and SD2, which reduce the
sound level of the blower.