Dr
ä
ger Medizintechnik GmbH
Alarm Messages Savina Version 2.0
volume controlled: Pmin = PEEPs5 mbar
pressure-controlled: Pmin = 0.5 * (PEEPsPinsp)
The minimum time is determined by:
Tmin = min (
¼
* Tinsp, measured, 100 ms)
Tinsp, measured, is derived from the phase information of the ventilation control.
1.
Detection 1 is activated only at PEEP
setting
≥
3 mbar.
2.
After cold start and 15 s afterwards.
3.
During standby and 15 s afterwards.
4.
During a detected disconnection and 15 s afterwards.
1.
Generation of event "Airway pressure low" (warning "!!! Airway pressure low"). The wa
is delayed in relation to the event by the adjustable time Tdisconnect.
1.
Detection 1: If the summation memory has been deleted.