Error List Savina Version 2.0
60.0200
Calculated airway pressure (Y-Paw):
Invalid measured value.
−
Sequence error, see error codes
60.0000 through 60.0012.
60.0300
Calculated inspiratory airway
pressures of FP and MP deviate too
much.
−
Sequence error, see error codes
60.0000 through 60.0012.
60.0310
Deviation between inspiratory airway
pressure and flow-corrected expiratory
airway pressure is too large.
−
Sequence error, see error codes
60.0000 through 60.0012.
61.0000
Sensor open wire.
−
Check connection between Spirolog
sensor and Control PCB.
−
Replace Spirolog sensor.
61.0001
Measured value is outside
specification.
−
Replace Spirolog sensor.
−
Check connection between Spirolog
sensor and Control PCB.