Dr
ä
ger Medizintechnik GmbH
Repair Procedure Savina Version 2.0
The power pack delivers the following supply voltages:
The +48 V are fed directly from the power pack to the Motor Commutator PCB. The remaini
voltages are fed to the Control PCB and from there to the other PCBs and component devices.
The voltages can be measured on the Control PCB (except the +48 V). The operator control p
must be opened to access the measuring points (see "
Opening the operator control panel
voltages are also displayed in Service Mode, chapter "
Test Step 4
".
+5 V
—
15 V
+15 V
+24 V
+48 V