![Dräger Savina Скачать руководство пользователя страница 108](http://html.mh-extra.com/html/drager/savina/savina_service-manual_2526836108.webp)
F
o
r
i
n
t
e
r
n
a
l
u
s
e
o
n
l
y
.
C
o
p
y
r
i
g
h
t
r
e
s
e
r
v
e
d
.
R
5
6
6
4
9
0
0
F
L
1
.
f
m
2
7
.
0
8
.
0
1
A
B
C
Dr
ä
ger Medizintechnik GmbH
Error List Savina Version 2.0
The error code is divided into a 2-digit decimal number and a 4-digit decimal number. The 2-digit
decimal number indicates which component/assembly is affected. The 4-digit decimal number is a
more detailed specification for the software as to the malfunction suffered by the
component/assembly concerned.
Example:
The "42" indicates that the error relates to the blower. The "0000" indicates that the rotation speed
is too low. The "malfunction description" is not always described in plain text, since it in some
cases involves names of software routines. "Possible remedy" shows the possible defective
assembly or a reference to other remedies.
42.0000
Rotation speed too low.
Replace the blower.