Discard Value Settings
Define the physiological range assigning min/max thresholds to the following
parameters:
Parameter
Description
Default Min
Default Max
VT (mL)
Expiratory breath volume 200
--
VO2 (mL/min)
Oxygen consumption
50
1000
RF (1/min)
Respiratory frequency
2
50
RQ
Respiratory quotient
0.5
1.5
BTPS
Enable/Disable BTPS correction (default=yes)
IntelliBridge
Enable/Disable the communication to a Multiparameter Monitor through the
IntelliBridge protocol (For for further details refer to “IntelliBridge Interface”).
Available on Q-NRG+ only.
Start RMR Avg (min)
Set the time when REE averaging automatically starts (default=5min)
Stop RMR Avg (min).
Set the time when REE averaging automatically stops (default=15min)
Min Canopy Dilution FeCO2(%)
Minimum FeCO2 triggering warning message (default=0.7%)
Max Canopy Dilution FeCO2(%)
Maximum FeCO2 triggering warning message (default=1.5%)
Substrate Metabolism (UM)
Select the unit of measurement for substrate visualization (%, kcal/day, g/day)
Mask Flowmeter Temp (°C)
Temperature at flowmeter during mask test (°C)
Printout
Printout menu allows configuration of facility or organization details to be printed in the header of a test report.
Field
Description
Print Detailed Steps
When enabled all tabular results recorded during the test are shown in a separate
page step by step
Header Line 1/2/3
Use these fields to add the information you wish to be printed on reports
Predicteds
Predicteds menu allows the selection of preferred predicted set from the following list:
Predicteds
Reference
ESPEN (25kcal/kg/day)
[ESPEN Guidelines on Enteral Nutrition: Intensive care, Clinical Nutrition (2006) 25,
210–223]
Harris Benedict (default)
[Harris J.A. BF: A biometric study of basal metabolism in Man. Washington DC,
Carnegie Institute of Washington, 1919]
Mifflin
[A new predictive equation for resting energy expenditure in healthy individuals,
Am J Clin Nutr 1990;51:241-7]
Owen
[Owen O.E., Kavle E., Owen R.S., Polansky M., Caprio S., Mozzoli M.A., Kendrick Z.V.,
Bushman M.C., Boden G.: A reappraisal of caloric requirements in healthy women.
Am. J. Clin. Nutr., 44, 1–19, 1986]
Содержание Q-NRG
Страница 1: ...Q NRG User Manual...
Страница 2: ......
Страница 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 33: ...31 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 2 Installation COSMED srl Chapter 2 Installation...
Страница 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 87: ...85 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 5 Maintenance COSMED srl Chapter 5 Maintenance...
Страница 109: ...107 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 6 Troubleshooting COSMED srl Chapter 6 Troubleshooting...
Страница 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Страница 122: ...120 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl CE Declaration of Conformity...
Страница 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Страница 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 151: ......