REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 3 - Q-NRG Operation
COSMED srl
Calibration Menu
Calibration functions are accessible by tapping on the
Calibration
tile from the Home page. Refer to the diagram
below to access the required calibration function.
Calibrations
Gas Analyzers
(Cylinder)
Ventilator
Mode
Canopy/Face
Mask Mode
Flowmeters
Pneumotach
Turbine
Blower
Gas Analyzers
(Room Air)
Ventilator
Mode
Canopy/Face
Mask Mode
Calibration
Factors
References
Results
Pneumotach (VENTILATOR Mode)
The pneumotach (Flow-REE) is a “variable orifice”
flowmeter, used during REE measurement with
ventilated subjects
The pressure drop across the transparent flexible
plastic screen is directly proportional to the amount of
air flow passing through the flowmeter.
To achieve the maximal possible accuracy a rapid
flowmeter calibration prior to each REE measurement
is mandatory.
A
Clear tube
B
Blue tube
The device will prompt the user to perform pneumotach calibration when starting a measurement in VENTILATOR
Mode. The same calibration procedure can be launched from the calibration menu. Flow-REE will be calibrated
against the flow of air generated by the internal Canopy blower, monitored by a pre-calibrated internal turbine
flowmeter, to generate a known amount of inhalation flow that will be compared with the one measured by the
single use pneumotach to calculate the ideal correction factor.
Содержание Q-NRG
Страница 1: ...Q NRG User Manual...
Страница 2: ......
Страница 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 33: ...31 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 2 Installation COSMED srl Chapter 2 Installation...
Страница 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 87: ...85 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 5 Maintenance COSMED srl Chapter 5 Maintenance...
Страница 109: ...107 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 6 Troubleshooting COSMED srl Chapter 6 Troubleshooting...
Страница 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Страница 122: ...120 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl CE Declaration of Conformity...
Страница 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Страница 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 151: ......