REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 3 - Q-NRG Operation
COSMED srl
Test Overview
The following process shows an overview of the step-by-step procedure to conduct a test.
#
Step
Description
1
Required Items
Collect the necessary items according to specific test mode
2
Warm-up and Calibrations
Follow the recommended warm-up time
Perform periodic calibrations when needed
3
Data entry and Test Selection
Select existing patient in the archive or insert a new one
Select the Test Mode
4
Test Set-up
Set-up device and accessories according to the on-screen images while the device
automatically calibrates
When ready select
Start
5
Real time execution
Check the parameters, when ready select
Start Recording
Supervise the test execution and adjust settings if necessary
For maximum test duration refer to “Appendix”
6
End of test
Disconnect the patient from the Q-NRG ensuring patient comfort and safety
Select
Stop Recording
to store the test
7
Edit Results and Save Changes
Select the averaging interval
Review and edit the test results and patient’s info (if necessary)
8
Export Results
Export in CSV or PDF format (USB port)
Export to OMNIA Software (Optional)
9
Dispose and Cleaning
Dispose non reusable items
Clean and disinfect reusable items according to the instructions in this manual
Operating by AC Power or Battery
Q-NRG can be powered by the rechargeable battery or plugged into the wall outlet. Prior to starting a
measurement using battery power, make sure that the residual capacity displayed on the LCD is enough to start a
test (at least 15% for a test lasting 10 minutes, depending also on the battery’s age).
Note:
Q-NRG can also be connected to external power during a test.
Warming-up Time
The sensors (O2, CO2, pressure, etc.) are sensitive to temperature changes. The temperature of an electrically
powered device may increase at the start of operation. Q-NRG has been designed to compensate for sensor thermal
drift using technological solutions including periodic auto-calibration (undetectable to the user). This allows an REE
test to begin after only 5 minutes of warm-up.
Содержание Q-NRG
Страница 1: ...Q NRG User Manual...
Страница 2: ......
Страница 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 33: ...31 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 2 Installation COSMED srl Chapter 2 Installation...
Страница 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 87: ...85 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 5 Maintenance COSMED srl Chapter 5 Maintenance...
Страница 109: ...107 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 6 Troubleshooting COSMED srl Chapter 6 Troubleshooting...
Страница 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Страница 122: ...120 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl CE Declaration of Conformity...
Страница 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Страница 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 151: ......