
REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 5 - Maintenance
COSMED srl
Mask
Level of Reprocessing
High-level disinfection
When
Immediately after use
Pre-treatment
Rinse in water at 22-40°C (72-104°F)
Manual Cleaning
Soak for 3 minutes in an enzymatic solution (e.g. Cidezyme) at room temperature (>22°C,
72°F). Using a soft bristle brush, scrub the item submerged in the cleaning solution during the
3 minutes. Wipe while rinsing for 5 minutes in water. Inspect all surfaces and item features to
ensure that it is visibly clean, repeat Manual Cleaning if not.
Disinfection
Soak the component in disinfectant (e.g. Cidex OPA Solution) at room temperature (>22°C, 72°F)
for 12 minutes. Rinse three (3) times in a volume of room temperature (>22°C, 72°F) water,
large enough to completely submerge the part, for at least 3-minutes for each rinsing cycle.
Drying
Dry the parts immediately after the disinfection steps by using filtered air (oil-free, low-germ
and low-particle) to blow them off/out or dab with a lint free cloth and let air dry. Inspect all
surfaces and item features to ensure that they are visibly dry, repeat Drying if not.
Note
Disassemble all parts before cleaning.
Material Facts
Silicone rubber (blue), Polycarbonate clear plastic. Adapter: ABS plastic
Head cap
Level of Reprocessing
Low-level disinfection
When
Immediately after use
Pre-treatment
Rinse in water at 22-40°C (72-104°F)
Manual Cleaning
Soak for 3 minutes in an enzymatic solution (e.g. Cidezyme) at room temperature (>22°C,
72°F). Using a soft bristle brush, scrub the item submerged in the cleaning solution during the
3 minutes. Wipe while rinsing for 5 minutes in water. Inspect all surfaces and item features to
ensure that it is visibly clean, repeat Manual Cleaning if not.
Disinfection
Repeat Manual Cleaning/Rinsing
Drying
Dry the parts immediately after the disinfection steps by using filtered air (oil-free, low-germ
and low-particle) to blow them off/out or dab with a lint free cloth and let air dry. Inspect all
surfaces and item features to ensure that they are visibly dry, repeat Drying if not.
Note
Do not iron the head cap assembly
Material Facts
Headgear: Polyurethane Foam, Nylon UBL Gray and Nylon Fabric Red. Headgear Hook: Nylon.
Headgear Strap Clips (4): Polypropylene.
Содержание Q-NRG
Страница 1: ...Q NRG User Manual...
Страница 2: ......
Страница 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 33: ...31 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 2 Installation COSMED srl Chapter 2 Installation...
Страница 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 87: ...85 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 5 Maintenance COSMED srl Chapter 5 Maintenance...
Страница 109: ...107 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 6 Troubleshooting COSMED srl Chapter 6 Troubleshooting...
Страница 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Страница 122: ...120 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl CE Declaration of Conformity...
Страница 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Страница 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 151: ......