REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 6 - Troubleshooting
COSMED srl
Error#104
Cylinder gas not detected
No calibration gas flowing into the
analyzers when expected
Review calibration set-up:
Cylinder connected
Cylinder valve Open
Output pressure 5-6bar (70-90psi)
Pressure regulator ON/OFF valve Open
Cylinder composition properly specified
in Calibration/Cal factors/References
Error#105
Environmental CO2 too high
Room air CO2 content higher than
maximum admissible (0.25%)
Make sure room is properly aerated
before performing calibration and test
Error#106
O2 sensor exhausted (Maintenance
required)
O2 output signal is lower than
minimum acceptable threshold due to
sensor aging
Replace the O2 sensor according to
maintenance instructions
Error#107
CO2 sensor not ready.
Unable to complete calibration.
CO2 sensor has not reached operative
temperature
Wait for a few minutes and try again
Failure#108
Calibration flow not detected.
Unable to complete calibration.
Internal Turbine Flowmeter does not
detect any flow although the internal
blower is properly functioning
Contact Service specifying error
message #
Error#109
Calibration flow too low
During pneumotach calibration,
internal turbine flow is too low
Calibrate internal flowmeter according
to instructions (Internal Turbine
Flowmeter - Blower Calibration) and
try again
Error#110
Calibration flow too high
During pneumotach calibration,
internal turbine flow is too high
Calibrate internal flowmeter according
to instructions (Internal Turbine
Flowmeter - Blower Calibration) and
try again
Error#111
Reversed flow detected. Swap
Flowmeter connections
Detected flow during Pneumotach
calibration is negative
Reverse Pneumotach connection
matching colour coding
Error#112
PNT flow not detected, check the
connections
Measured flow during Pneumotach
calibration is zero, regardless of proper
functioning of internal blower and
Turbine Flowmeter
Be sure the Pneumotach is correctly
connected to the device
Failure#115
Blower flow not detected
Internal Blower not functioning
Contact Service
Failure#116
Calibration Fail
Generic fail related to the calibration
procedure
Repeat the calibration; if the problem is
not solved, contact service
Failure#306
Blower Ventilation not detected,
remove the Canopy hood from
patient.
During Canopy test dilution flow is
not detected due to blower or internal
flowmeter failure
Remove Canopy hood as soon as possi-
ble and contact technical service
Содержание Q-NRG
Страница 1: ...Q NRG User Manual...
Страница 2: ......
Страница 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 33: ...31 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 2 Installation COSMED srl Chapter 2 Installation...
Страница 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 87: ...85 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 5 Maintenance COSMED srl Chapter 5 Maintenance...
Страница 109: ...107 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 6 Troubleshooting COSMED srl Chapter 6 Troubleshooting...
Страница 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Страница 122: ...120 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl CE Declaration of Conformity...
Страница 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Страница 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 151: ......