REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 5 - Maintenance
COSMED srl
Canopy hood
Level of Reprocessing
High-level disinfection
When
Immediately after use
Pre-treatment
Wipe soil with a moist sponge or towel
Manual Cleaning
Wipe the canopy surface with a soft cloth sprayed with AHP® cleaning solution (e.g. Oxyvir® Tb
RTU) for 30 seconds. Dispose the soft cloth.
Disinfection
Wipe the canopy surface with a soft cloth sprayed with AHP® cleaning solution (e.g. Oxyvir®
Tb RTU) for 60 seconds.Dispose the soft cloth. Wipe the device surface with another soft cloth
moistened with water for 30 seconds. Dispose the soft cloth.
Drying
Air dry
Note
Disassemble all parts before cleaning. DO NOT use any alcohol products, this will damage
material surface.
Material Facts
PMMA canopy, ABS connectors
Canopy hose
Level of Reprocessing
High-level disinfection
When
It must be cleaned monthly or replaced.
Pre-treatment
Thoroughly brush the inside (at the ends) and outside, wipe and rinse in water at 22-40°C (72-
104°F)
Manual Cleaning
Soak for 3 minutes in an enzymatic solution (e.g. Cidezyme) at room temperature (>22°C, 72°F).
During the 3 minutes, using a soft bristle brush, scrub the item submerged in the cleaning solu-
tion.Wipe while rinsing for 5 minutes in water. Inspect all surfaces and item features to ensure
that it is visibly clean, repeat Manual Cleaning if not.
Disinfection
Soak the component in disinfectant (e.g. Cidex OPA Solution) at room temperature (>22°C, 72°F)
for 12 minutes. Rinse three (3) times in a volume of room temperature (>22°C, 72°F) water, large
enough to completely submerge the parts, for at least 3-minutes for each rinsing cycle.
Drying
Dry the parts immediately after the disinfection steps by using filtered air (oil-free, low-germ
and low-particle) to blow them off/out or dab with a lint free cloth and let air dry. Inspect all
surfaces and item features to ensure that they are visibly dry, repeat Drying if not.
Note
Do not disassemble adapters, doing so will damage hose. During test, it must be used in conjunc-
tion with a COSMED antibacterial filter REF A-182-300-004.
Material Facts
ABS adapter, EVA tube
Содержание Q-NRG
Страница 1: ...Q NRG User Manual...
Страница 2: ......
Страница 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 33: ...31 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 2 Installation COSMED srl Chapter 2 Installation...
Страница 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 87: ...85 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 5 Maintenance COSMED srl Chapter 5 Maintenance...
Страница 109: ...107 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 6 Troubleshooting COSMED srl Chapter 6 Troubleshooting...
Страница 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Страница 122: ...120 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl CE Declaration of Conformity...
Страница 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Страница 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 151: ......