REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 6 - Troubleshooting
COSMED srl
Test Notification
Canopy dilution not detected,
check hood connections
During Canopy expired CO2
not detected. Probable wrong
test set-up
Remove Canopy hood as soon as possible, careful-
ly re-check all connections and try again
Test Notification
Warm up time less than 5
minutes, do you want to run
measurement anyway?
Test execution requires 5
minutes warm-up
Wait for 5 minutes before performing test
Test Notification
Current test results appear
technically questionable:
carefully evaluate if suitable for
clinical use
At the end of test or during
test edit, results are remark-
ably outside physiological
range
Carefully evaluate if data are suitable for clinical
use
Test Notification
Test duration exceeding 30
minutes limit. Potential water
condensation within sampling
line
Ventilator test duration
exceed 30 minutes
Terminate the test to avoid possible water
condensation within the sampling line
Содержание Q-NRG
Страница 1: ...Q NRG User Manual...
Страница 2: ......
Страница 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 33: ...31 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 2 Installation COSMED srl Chapter 2 Installation...
Страница 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 87: ...85 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 5 Maintenance COSMED srl Chapter 5 Maintenance...
Страница 109: ...107 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 6 Troubleshooting COSMED srl Chapter 6 Troubleshooting...
Страница 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Страница 122: ...120 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl CE Declaration of Conformity...
Страница 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Страница 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 151: ......