REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Chapter 3 - Q-NRG Operation
COSMED srl
Operating Q-NRG in VENTILATOR Mode
1. Real-time data are displayed on-screen, measurements are displayed after approximately 90 seconds.
2. Data are stored automatically after 2 minutes (or as defined in Settings/Advanced), tap
Start Recording
to
store data immediately.
Note:
Measurements will be shown after approximately 90 seconds from the beginning of acquisition.
Dashed data displayed indicates that valid values are not yet available or do not apply.
During test execution user interacts via the following commands:
Command
Function
Start Recording
Begin storage of data, test time will reset
Stop Recording
End test storing all data
Info
Review/edit additional test info (UN, Fasting Period, Resting Period, Test Position,
Body Temperature, Agitation Level, Sedation Level, Test Purpose, Operator Notes
and Operator)
MARKER
Insert a marker
Read FiO2
Trigger a new FiO2 reading as soon as possible
Abort
End test discarding all data
Actual
Shows real-time values (updated every 30 seconds)
AVG
Shows averaged results within the specified interval (manually adjustable using
the touch display)
Data
“Data” tab shows values in tabular format (updated every 30 seconds) During test,
main parameters are updated every 30 seconds in both numerical and graphical
formats
Содержание Q-NRG
Страница 1: ...Q NRG User Manual...
Страница 2: ......
Страница 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 33: ...31 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 2 Installation COSMED srl Chapter 2 Installation...
Страница 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 87: ...85 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 5 Maintenance COSMED srl Chapter 5 Maintenance...
Страница 109: ...107 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 6 Troubleshooting COSMED srl Chapter 6 Troubleshooting...
Страница 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Страница 122: ...120 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl CE Declaration of Conformity...
Страница 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Страница 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 151: ......