REF C04717-02-91 Q-NRG User Manual, Edition 4.2
Appendix
COSMED srl
Other information
Contraindications for Indirect Calorimetry testing
Refer to guidelines reported in the References of this appendix and specific country regulations.
Environmental condition of use
COSMED units should not be operated near explosive substances. Device should not be installed near electrical
or magnetic devices such as x-ray equipment, transformers or power lines. These devices could create electrical
interferences when performing testing procedures. COSMED devices are not AP or APG units (according to IEC
60601-1) and should never be operated in the presence of flammable anaesthetic mixtures.
•
Temperature range: 10°C (50°F) to 35°C (95°F).
•
Relative humidity range: 5% to 93% (non-condensing).
•
Atmospheric Pressure range: 700hPa to 1060hPa (up to 3011 m / 9875 ft).
•
Avoid operating device in the presence of noxious fumes or in dusty environments.
•
Do not place Q-NRG near heat sources.
•
Adequate ventilation should be maintained in the room the testing is performed.
Manufacturer’s Information
You may contact the manufacturer directly at the following address for information:
COSMED Srl
Via dei Piani di Monte Savello, 37
00041 - Albano Laziale (RM) - ITALY
Voice:+39 (06) 9315492 Fax:+39 (06) 9314580
Email:
Internet:
Worldwide COSMED offices are listed in corporate web page
https://www.cosmed.com/en/contact-us
Complaints, Feedback and Suggestions
If you have any complaints, feedback or suggestions you may inform us at
Privacy Information
Refer to Privacy disclaimer reported in corporate web site:
https://www.cosmed.com/en/company/privacy
Disposing of electrical equipment
The device cannot be disposed as unsorted municipal waste. Electronic equipment must be collected separate-
ly according to the European Directive 2012/19/EU. Otherwise it can cause dangerous consequences for the
environment and human health.
The crossed-out wheeled bin means that the product must be taken to a separate collection when you wish to
dispose of it.
Содержание Q-NRG
Страница 1: ...Q NRG User Manual...
Страница 2: ......
Страница 32: ...30 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 33: ...31 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 2 Installation COSMED srl Chapter 2 Installation...
Страница 40: ...38 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 86: ...84 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 COSMED srl...
Страница 87: ...85 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 5 Maintenance COSMED srl Chapter 5 Maintenance...
Страница 109: ...107 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Chapter 6 Troubleshooting COSMED srl Chapter 6 Troubleshooting...
Страница 121: ...119 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Appendix...
Страница 122: ...120 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl CE Declaration of Conformity...
Страница 148: ...146 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl Notes...
Страница 149: ...147 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 150: ...148 148 REF C04717 02 91 Q NRG User Manual Edition 4 2 Appendix COSMED srl...
Страница 151: ......