Subassembly Removal and Replacement
5
Required Tools:
3 mm hex wrench.
5 mm hex wrench.
Preparation:
1. Turn the sterilizer OFF.
2. Remove the front top panel, top panel, front bottom panel, and right
side panel.
Removal:
1. Unplug cable P6.
2. Remove the two screws that attach the diagonal bracket in front of the
vacuum subsystem to the chassis.
3. Remove two screws that secure the vacuum subsystem platform.
4. Disconnect the Twinax feedthrough cable from the bottom of the
feedthrough.
5. Disconnect the vacuum hose from the chamber.
6. Unplug two cables: P112 and P113.
WARNING! THE PUMP AND PLATFORM ARE VERY HEAVY
(29.5 KG). USE CARE WHEN REMOVING
7. Grasp the front handle of the pump platform and pull the platform
straight out. It will slide forward on guides.
Use caution, the pump
and platform weighs about 29.5 kg (65 lbs).
Replacement:
Replacement is the reverse of the removal steps.