7
System Calibration and Verification
On Calibration Tests (shown at left) touch
Temperature Calibration.
“Set the Bath Temperature to 35.0
˚
C” is
displayed.
Figure 49. Temperature Calibration Showing Virtual Keypad.
2. Select each heater circuit you wish to calibrate (you can also select
ALL). Press
Start
to turn off the heater and begin the calibration
procedure.
3. Set the bath temperature to 35.0º C and place the thermistors in the
bath.
4. When all thermistors selected for calibration are within 3º C of the
setpoint, the virtual keypad and the Actual Bath Temperature entry
fields pop-up to allow entry of the actual temperature of the bath.
Press
Enter
to accept and save the value.
5. “Set Bath Temperature to 50.0º C” is displayed. Set the bath
temperature to 50.0º C, entering actual bath temperature when it is
stable. Repeat step 4.
120 STERRAD
®
NX
™
Service Guide
Содержание sterrad NX
Страница 1: ...STERRAD NX Sterilization System Service Troubleshooting Guide...
Страница 10: ......
Страница 16: ......
Страница 21: ...Functional Description 3 Figure 1 Software Control Diagram STERRAD NX Service Guide 21...
Страница 38: ......
Страница 100: ......
Страница 124: ......