Diagnostic Tests and Error Messages
4
If you see this message . . .
Perform this action.
LOW PLASMA POWER
Run diagnostics to determine the nature of the
power problem. If the diagnostics fail, replace
the LFPS II power supply.
MAIN FAN FAULT
Inspect the fan. Replace if needed.
MEMORY CARD FULL
Store the cycle record information from the
PCMCIA card on an external PC.
PLEASE REMOVE CASSETTE AND
VERIFY CASSETTE ORIENTATION or,
PLEASE REMOVE CASSETTE AND
VERIFY CASSETTE TYPE
Inspect the cassette. Make sure the barcode is
clean and legible. Orient the cassette and insert
it. If problem persists, insert a new cassette. If
problem persists, run diagnostics. Adjust or
replace the delivery subsystem.
PRESSURE OUT OF RANGE [HIGH] or,
PRESSURE OUT OF RANGE [LOW]
Verify load is dry and not outgassing. Start a
new cycle. If the pressure cannot be maintained
during vacuum, run diagnostics. Replace failed
component.
PRESSURE SENSOR BELOW
ATMOSPHERE, RUN DIAGNOSTICS
Run diagnostics to determine the nature of the
fault detected. Replace failed component.
PRINTER IS OUT OF PAPER. PLEASE
LOAD A NEW ROLL
Insert a new roll of paper. If problem persists,
replace printer assembly.
POWER SUPPLY OUT OF RANGE
The power supply voltage is 20% higher or
lower than needed. Run diagnostics to
determine the cause and replace the failed
component.
PRESSURE OUT OF RANGE (Low)
SYSTEM FAILURE - CALL ASP or,
SYSTEM PROBLEM – CALL ASP
Run system diagnostics to determine the cause of
the problem. Reload software or install new
software if needed.
TEMPERATURE HAS NOT RISEN
If the sterilizer door is open, close the door and
wait 30 minutes to bring the system to operating
temperature. If the problem persists, run
diagnostics to determine if a heater or sensor has
failed. Replace failed component.
UNABLE TO LIGHT PLASMA. USE
EXTRA CAUTION WHEN HANDLING
THIS LOAD
The plasma did not light during canceled cycle.
Run system diagnostics to determine the cause of
the problem. Replace the failed component.
UNABLE TO OPEN ONE SECOND DATA
FILE
Install a new PCMCIA compact flash memory
card.
STERRAD
®
NX
™
Service Guide
45