Subassembly Removal and Replacement
Vacuum Control Valve (28-51216)
Figure 34. Vacuum Control Valve
4
Note
: You may need to perform this procedure first in
order to remove the vacuum pump.
Required Tools:
None.
Preparation:
1. Remove the front top panel, top panel, front bottom panel, and
right side panel.
2. Disconnect the KF-25 clamp from the chamber.
3. Remove the vacuum pump assembly.
Removal:
1. Loosen the Ultratorr fitting attaching the vacuum control valve
to the chamber.
2. Unplug cable P112.
3. Remove the tubing clamp and remove the valve.
Replacement:
Replacement is the reverse of the removal steps.
94 STERRAD
®
NX
™
Service Guide