Subassembly Removal and Replacement
5
Hydrogen Peroxide Detector Assembly (04-52108)
The hydrogen peroxide detector assembly includes the detector optical
window.
O-Ring
Inside
Optical
Window
Inside
Figure 30. Detector Assembly
CAUTION: Safe parts handling. Do not touch the surface of the
optical window with your bare hands. Oil and contaminants
from your skin will damage the window and cause premature
failure
.
Required Tools:
3 mm hex wrench.
Preparation:
1. Turn the sterilizer OFF
2. Remove the front bottom panel.
Removal:
1. Unplug cable P16.
2. Unscrew the detector assembly from the chamber fitting.
3.
Do not touch the optical window with your bare hands.
Remove the O-ring and the optical window.
Replacement:
1.
Do not touch the optical window with your bare hands
.
Install a new optical window and O-ring in the detector (if
necessary).
2. Clean the inside of the chamber fitting.
3. Lightly coat a new O-ring with approved vacuum grease and
place the O-ring in the chamber fitting.
4. Screw the detector onto the fitting and hand-tighten.
5. Reconnect cable P16.
STERRAD
®
NX
™
Service Guide
89