Subassembly Removal and Replacement
Removal:
1. Unplug cable P17.
2. Remove the 4 screws attaching the printer module to the chassis.
Support the printer module as the last screw is removed.
3. Pull the printer module out and to the right to remove (avoid
contact with the hydrogen peroxide monitor detector).
Replacement:
Replacement is the reverse of the removal steps.
98 STERRAD
®
NX
™
Service Guide