4
Diagnostic Tests and Error Messages
Troubleshooting
Most sterilizer operating problems are accompanied by an error message.
These messages are useful in determining the source of the problem. In
many cases you can take remedial actions to correct the source of the
problem and thereby return the sterilizer to normal operation. In other
cases, the problem is caused by a component failure that requires
adjustment or replacement.
Error Messages
If an error occurs during operation, the sterilizer may allow you to run a
diagnostic test immediately. When prompted to do so, touch
Confirm
to
perform the diagnostic test. A diagnostic test displays and prints a
diagnostic message when the test is completed. Some diagnostic
messages indicate conditions that you may be able to remedy.
If you see this message . . .
Perform this action.
12 VOLT SUPPLY OUT OF RANGE or,
15 VOLT SUPPLY OUT OF RANGE or,
24 VOLT SUPPLY OUT OF RANGE or,
3.3 VOLT SUPPLY OUT OF RANGE or,
5 VOLT SUPPLY OUT OF RANGE
Run the diagnostics to determine if the power
supply is out of tolerance. If the diagnostics fail,
replace the power supply.
AUTO ADJUSTMENT FAILED
A fault was detected during the hydrogen
peroxide auto-adjustment.
BARCODE DATA ERROR, EJECTING
CASSETTE
or
BARCODE NOT DETECTED, EJECTING
Inspect the cassette to determine if the barcode is
valid. Make sure the barcode is clean and
legible. Insert the cassette again or try a new
cassette. If the problem persists, run the
diagnostics. Adjust or replace the barcode
sensor.
CANNOT DISPOSE CASSETTE, RUN
DIAGNOSTICS
Never put your hand inside the machine to move
a cassette. Select “Dispose Cassette” to move the
cassette into the disposal drawer. If the problem
persists, run diagnostics. If diagnostics fail,
adjust or replace the delivery subsystem.
CANNOT EJECT CASSETTE, RUN
DIAGNOSTICS
Run diagnostics. If diagnostics fail, adjust or
replace the delivery subsystem.
42 STERRAD
®
NX
™
Service Guide
Содержание sterrad NX
Страница 1: ...STERRAD NX Sterilization System Service Troubleshooting Guide...
Страница 10: ......
Страница 16: ......
Страница 21: ...Functional Description 3 Figure 1 Software Control Diagram STERRAD NX Service Guide 21...
Страница 38: ......
Страница 100: ......
Страница 124: ......