Chapter 7.
System Calibration and
Verification
Overview
The STERRAD
®
NX
™
Sterilizer has onboard calibration tools to make it
easy to calibrate the various components of the system. As of this
publication, calibration is still being researched and the procedures
presented here are subject to change. Please contact ASP Technical
Service prior to performing any calibration procedure.
Service Action Table
Use the system verification procedures to verify sterilizer performance
whenever the service actions listed in the following table are completed.
Service Action Performed
System Verification Test Requirements
PM
All tests.
All Service
Vacuum plasma test, plasma power measurement,
temperature verification, diagnostics other tests and run
standard cycle.
Replace vacuum subsystem
components
Vacuum plasma test, leak test, plasma power
measurement, temperature verification, diagnostics
other tests and run standard cycle.
Replace plasma subsystem
components
Vacuum plasma test, leak test, plasma power
measurement, temperature verification, diagnostics
other tests and run standard cycle.