Subassembly Removal and Replacement
5
Required Tools:
3 mm hex wrench.
Preparation:
1. Turn the sterilizer OFF
2. Remove the front top panel and the top panel.
3. Remove the UCB/interface side cover.
Removal:
1. Unplug two cables: P6 and P44.
2. Move the outside edge of the power supply cover to expose the
screw head.
3. Remove four screws that attach the DC power supply to the
chassis.
4. Lift the DC power supply up to remove.
Replacement:
Replacement is the reverse of the removal steps.