System Calibration and Verification
7
Temperature Calibration Using A
Temperature Bath or Precision
Resistors
Temperature ranges are calibrated or verified in this procedure. If
precision resistors are used, follow the same process as shown below,
except remove the thermistors and substitute a resistor for each of the
following thermistors: 35.0º, 50.0º, 60.0º, 70.0º, and 85.0º C.
To perform the calibration using the temperature bath, do the following:
1. Log in as Service. Navigate to the Temperature Calibration display
by touching the buttons in the displays shown in the next page.
On Additional Utilities (shown at left) touch
Service Functions
On Service Functions (shown at left) touch
Calibration
STERRAD
®
NX
™
Service Guide
119