
5
Subassembly Removal and Replacement
Alternating Current (AC) Distribution Assembly (02-51959)
The AC distribution assembly contains a voltage configuration board
which
must be set to match the input voltage for the sterilizer
. A
replacement AC distribution assembly may not be set for the correct
voltage. When using a new AC distribution assembly, verify that the
configuration board is set for the correct input voltage. If the
configuration board is incorrect, unplug the board from the connector,
turn the board upside down, verify the label, and plug it back into the
connector.
Figure 20. AC Distribution Assembly
72 STERRAD
®
NX
™
Service Guide
Содержание sterrad NX
Страница 1: ...STERRAD NX Sterilization System Service Troubleshooting Guide...
Страница 10: ......
Страница 16: ......
Страница 21: ...Functional Description 3 Figure 1 Software Control Diagram STERRAD NX Service Guide 21...
Страница 38: ......
Страница 100: ......
Страница 124: ......