Subassembly Removal and Replacement
5
Top Panel (33-51196)
Screw
Screw
Figure 9. Top Panel
Required Tools:
3 mm hex wrench.
Preparation:
1. Turn the sterilizer OFF.
2. Remove the front top panel.
Removal:
1. Remove the two screws on the front bottom left and right sides of
the top panel.
2. Lift the panel and pull it forward to remove.
Replacement:
Replacement is the reverse of the removal steps.
STERRAD
®
NX
™
Service Guide