Cassette Handling
STERRAD CASSETTES CONTAIN CONCENTRATED HYDROGEN
PEROXIDE, A STRONG OXIDIZER. CONCENTRATED
HYDROGEN PEROXIDE IS CORROSIVE TO SKIN, EYES, NOSE,
THROAT, LUNGS, AND GASTROINTESTINAL TRACT.
Direct
contact with the skin can cause severe irritation. If skin contact occurs,
immediately flush with large amounts of water. If symptoms are severe or
persist, consult a physician immediately. Direct contact with eyes can cause
irreversible tissue damage. If eye contact occurs, immediately flush with
large amounts of water and immediately consult a physician. Inhalation of
vapor or mist can cause severe irritation of lungs, throat, and nose. If
inhalation occurs, move to fresh air and consult a physician immediately.
Ingestion can produce corrosion that may be life threatening. If swallowed,
immediately drink plenty of water to dilute. Do not induce vomiting. Consult
a physician.
Do not remove the plastic wrapper from the cassette package if the indicator
strip is red. Red indicates that the cassette might have been damaged. Call
your ASP representative for credit.
Do not remove used cassettes from the cassette collection box. Dispose of
the sealed cassette collection box according to local waste regulations.
Cassettes with unused hydrogen peroxide are hazardous waste as defined by
the Environmental Protection Agency (USA) and should be disposed of
accordingly. I
f it is necessary to handle a used cassette, wear latex, PVC
(vinyl), or nitrile gloves. Do not touch gloves to face or eyes.
Empty or expired cassettes must be replaced prior to starting the cycle as
directed by a message on the sterilizer display. Cassettes with unused
hydrogen peroxide are hazardous waste as defined by the Environmental
Protection Agency (USA) and should be disposed of accordingly.
Personal Protection Equipment
Wear latex, PVC (vinyl), or nitrile gloves whenever handling a load after a cycle
cancellation. Hydrogen peroxide liquid may be present on the load or in the
chamber.
STERRAD
®
NX
™
Service Guide
13
Содержание sterrad NX
Страница 1: ...STERRAD NX Sterilization System Service Troubleshooting Guide...
Страница 10: ......
Страница 16: ......
Страница 21: ...Functional Description 3 Figure 1 Software Control Diagram STERRAD NX Service Guide 21...
Страница 38: ......
Страница 100: ......
Страница 124: ......