Subassembly Removal and Replacement
Door Cover (33-51993)
Screw
Screw
Screw
Screw
Figure 14. Door Cover
WARNING! RISK OF ELECTRIC SHOCK AND BURNS! THE
INTERIOR OF THE DOOR CONTAINS ELECTRICAL
COMPONENTS THAT MAY BE HOT AND MAY CAUSE AN
ELECTRIC SHOCK. TURN OFF THE STERILIZER BEFORE
REMOVING THE DOOR COVER
.
Required Tools:
3 mm hex wrench.
Preparation:
1. Turn the sterilizer OFF.
2. Wait until the door is cool to the touch before proceeding.
Removal:
1. Open the door.
2. Remove the four screws on the inside face of the door.
3. Hold the door handle and pull the door cover straight off the door
and remove.
Replacement:
Replacement is the reverse of the removal steps.