5
Subassembly Removal and Replacement
Catalytic Converter (25-05611)
Figure 32. Catalytic Converter.
This is a PM procedure.
Required Tools:
None.
Preparation:
1. Turn the sterilizer OFF.
2. Remove the front top panel, top panel, front bottom panel, and
right side panel.
3. Remove the vacuum pump assembly if needed.
Removal:
1. Unscrew the catalytic converter from the base and discard.
Replacement:
1. Align the catalytic converter with the base, screw the parts
together and hand-tighten.
92 STERRAD
®
NX
™
Service Guide