Subassembly Removal and Replacement
5
Door Lock Solenoid (42-52001) and Door Lock Sensor
(04-51990)
The door lock solenoid and the door lock sensor are located together in
the top module. The latch and the door close sensor are mounted on the
door and the chassis near the chamber. Removal and replacement
procedures for the latch are described in the “Chamber Module” section
of this chapter.
Screw
Screw
Figure 24. Door Lock Assembly
WARNING! HOT SURFACES! THE SOLENOID BODY IS HOT
WHEN IT IS ENERGIZED (WHEN THE DOOR IS
UNLOCKED). AVOID CONTACT WITH THE ENERGIZED
SOLENOID.
Required Tools:
3 mm hex wrench.
Preparation:
1. Turn the sterilizer OFF.
2. Remove the front top panel and the top panel.
Removal:
1. Unplug two cables: P114 and P115.
2. Remove the two screws that attach the solenoid and sensor bracket
to the chassis.
3. Lift the solenoid and bracket up to remove.
Replacement:
Replacement is the reverse of the removal steps. It is important that the
door lock arm is centered as closely as possibly with the chassis slot
and the door panel slot. This is to that when the solenoid swings it
does not interfere with the slot.
STERRAD
®
NX
™
Service Guide
79