Introduction
1
Each parameter is part of a control loop in which information flows as
input from one of many sensors to the computer, where it is processed
and triggers an output signal which flows from the computer to an output
device. Through this system of feedback signals (input), analysis, and
response signals (output), the computer controls the entire sterilization
process. If any process parameter falls outside allowable ranges, the
software will cancel the cycle.
4
Note:
The hydrogen peroxide monitor does not provide
feedback control to the sterilizer. However, the
monitor will cancel the cycle if the area under the
hydrogen peroxide concentration-time curve or
hydrogen peroxide rate constant does not meet
specifications.