5
Subassembly Removal and Replacement
Low Frequency Plasma System (LFPS II) (40-52154)
Screw
Screw
Screw
Screw
Figure 36. LFPS II
WARNING! ELECTRIC SHOCK HAZARD. HAZARDOUS
VOLTAGES ARE PRESENT INSIDE THE LFPS II. TURN
THE STERILIZER OFF BEFORE HANDLING THE LFPS II.
Required Tools:
5 mm hex wrench.
Preparation:
1. Remove the following panels: front top, top, front bottom, and
right side.
Removal:
1. Unplug cable P6.
2. Unscrew the Twinax connector on the cable to the feedthrough.
3. Remove 4 screws attaching the LFPS II to the chassis.
4. Lift the LFPS II up and forward to remove.
Replacement:
Replacement is the reverse of the removal steps.
96 STERRAD
®
NX
™
Service Guide