5
Subassembly Removal and Replacement
Vent Valve (28-51218)
HEPA Filter
Ultratorr
Fitting
Valve Body
Figure 17. Vent Valve
Required Tools:
Soft-jaw pliers.
Preparation:
1. Turn the sterilizer OFF.
2. Remove the front top panel and the top panel.
Removal:
1. Unplug cable P4.
2. Unscrew the HEPA filter from the valve body (the HEPA filter
can remain in place if needed).
3. Use soft-jaw pliers to unscrew the collar connecting the valve to
the Ultratorr fitting.
4. Pull the valve up to remove.
5. Remove the spacer and the O-ring from inside the Ultratorr fitting.
66 STERRAD
®
NX
™
Service Guide