7
System Calibration and Verification
Peroxide Calibration With Neutral
Density Filters
1. Reinstall all the panels except for the bottom front panel.
2. Navigate to the Service/Calibration/Neutral Density display. Press
Set Filter.
Figure 50. Neutral Density Filter Calibration.
3. Enter the absorbance values marked on each filter in the order shown
on the display. Press
SAVE
. Press
BACK
.
4. If the system is not ready, START will be grayed out. During this
time, the system is setting up the sterilizer to support the calibration
and setting the temperature of the condenser to 70
˚
C to ensure no
residual peroxide remains in the chamber when the calibration
process begins. This setup may take several minutes. When the
sterilizer is ready, START will become active. Press
START
to
begin the calibration process. Remove the filter blank.
5. Follow the calibration steps on the display. Press
OK
after you have
completed each step that requires you to provide action. The first
action is to remove the filter blank. Remove the filter blank and press
OK
to continue.
6. Proceed with the calibration by following all steps on the display.
When the calibration is complete, the following is displayed:
Reinstall filter blank
122 STERRAD
®
NX
™
Service Guide