Subassembly Removal and Replacement
5
Delivery Valve
Required Tools:
Small pick
Preparation:
1. Turn the sterilizer OFF.
2. Remove the front top panel and the top panel.
3. Remove the delivery system from the sterilizer.
Removal:
1. Unscrew and disconnect P128 from the delivery valve.
2. Using the small pick, remove the 2 gaskets from the inside of the
fitting.
Replacement:
Replacement is the reverse of the removal steps. Note that new gaskets
come with the valve.
Delivery Subsystem Inlet Filter (PM kit part 25-50703)
This is a PM procedure.
Required Tools:
3 mm hex wrench.
Preparation:
1. Open the cassette drawer and remove the box of disposed cassettes
if present.
2. Remove the front top panel and the top panel.
Removal:
1. Slide the carriage on its lead screw toward the rear of the sterilizer
to cover the piercing needles.
2. Unscrew and disconnect the inlet filter cartridge.
3. Connect a new inlet filter to the inlet fitting.
Replacement:
Replacement is the reverse of the removal steps.
Air Pump Tubes
This is a PM Procedure
Preparation:
Remove the needle shroud.
Removal:
1. Unscrew the air pump from the bracket.
2. Remove the two tubes on the air pump.
Replacement:
Replacement is the reverse of the removal steps.
STERRAD
®
NX
™
Service Guide
63