Subassembly Removal and Replacement
5
Circuit Breaker (57-52105)
Circuit
Breaker
Circuit
Breaker
Bracket
Figure 38. Circuit Breaker
WARNING! ELECTRIC SHOCK HAZARD. HAZARDOUS
VOLTAGES ARE PRESENT AT THE CIRCUIT BREAKER.
TURN THE STERILIZER OFF AND UNPLUG IT FROM THE
MAINS BEFORE SERVICING THE CIRCUIT BREAKER.
Required Tools:
None.
Preparation:
1. Turn the sterilizer OFF.
2. Unplug the sterilizer from the mains.
3. Remove the front bottom panel.
4. Remove the right side panel.
Removal:
1. Pull off four push-on type lead connections on the rear of the
circuit breaker. Note the color-coded location of each lead.
2. From the rear of the circuit breaker, note the orientation of the
circuit breaker, then squeeze together the two tabs on each side of
the circuit breaker.
3. Push the circuit breaker assembly out of its bracket toward the
front of the chassis as far as it will move.
4. From the front of the sterilizer, grasp the circuit breaker and pull
it straight out from its bracket to remove.
Replacement:
Replacement is the reverse of the removal steps. Make sure the circuit
breaker has the same orientation as it had when originally installed.